Network analyzer time domain
With the time domain Option, FieldFox com-
putes the inverse Fourier transform of the
frequency-domain data to display reflection
or transmission coefficients versus time.
Time domain gating can be used to remove
unwanted responses such as connector
mismatch or cable discontinuities, and the
results can be displayed in either time or
frequency domain. FieldFox’s time domain
function supports both low pass mode and
band pass mode, enabling users to measure
both broadband and frequency-selective
devices.
Network analysis
S11, S11 phase, Smith chart and polar
display are available with Option 303 (Net-
work analysis capability). To obtain S21or
insertion loss/gain, users need to order
Option 110 (transmission measurement),
in addition to Option 303.
For in-fixture measurements, use FieldFox’s
port extension or electrical delay capabil-
ity to easily extend the reference plane to
the device interface to provide accurate
measurements. You can use the electrical
delay capability to measure deviation from
linear phase by removing the linear portion
of the phase delay.
If you need to measure the magnitude and
phase of all four S-parameters, consider the
N9923A FieldFox RF VNA. Please refer to
Agilent FieldFox RF Vector Network Ana-
lyzer, literature part number 5990-5087EN.