November, 2004 PcCR 1417 ACL4 System Automatic Cassette Loading Version Installation Manual Catalog number: AT095047-01

Strona 2

3. Site Preparation 3.1 General Inform the client about the site preparation requirements prior to installing the PcCR 1417. Without these requirem

Strona 3 - WARNING

Figure 3: PcCr 1417 DimensionsElectrical Requirements 50 cm 80 cm 100 cm Voltage: 90 V–250 V AC Frequency: 50Hz–60Hz Fuses: Two fuses are located nex

Strona 4 - Laser Safety Instructions

be exposed to light only when they move from the cassette into the PcCR 1417 scanner. For optimal results, room light is recommended at less than 100

Strona 5 - Table of Contents

4. Installation This chapter explains the installation procedure of the PcCR 1417. The FSE who installs the PcCR 1417 shall follow the steps descri

Strona 6

4.2 On-Site System Configuration PC PC PC PCServer Figure 4: System Configuration PC – Make sure that the PC you are using meets the requireme

Strona 7 - List of Figures

Installation 4-3 4.3 Equipment Required For Complete Installation • PC (refer to “computer requirements” for details) • Cables: Power cable (not

Strona 8 - 1. Terms and Conventions

4.4 Installation process The installation process includes the following: 4.4.1 Connecting the cables 4.4.2 Installing drivers for PcCR 1417 4.4.

Strona 9 - 2. Packaging

3. Choose “Search for a suitable…”; then click “Next” 4. Leave only the “specify a location” check-box selected; then click “Next” Installati

Strona 10 - 3. Site Preparation

5. Insert the “scanner interface” CD in to the CD-Rom drive. If installation does not begin automatically, Click the “Browse” button. 6. A “Loca

Strona 11 - Electrical Requirements

8. Mark the “orex02.inf” file; then click “Open” 9. The Installation Wizard finds the driver for the PcCR 1417. Click “Next”. 10. Click “Yes”

Strona 12

OREX Ltd. is a market innovator in the field of portable Computed Radiography Technologies—designed and priced for low-volume clinic or field unit

Strona 13 - 4. Installation

4.4.3 Installing the “Scanner Interface” 1. Insert the “Scanner Interface” CD into the CD-Rom drive. If installation does not begin automatically,

Strona 14 - Interface

4. Click “Next”. 5. Choose “Install new filters” and click “Next”. The software installs. Installation 4-9

Strona 15 - Installation

6. Choose the option appropriate to the specific system. Click “Next”. Installation 4-10

Strona 16 - 4.4 Installation process

7. Click Next 8. Use the default options and Click “Next” 9. Click “Next” Installation 4-11

Strona 17

10. When all tasks are completed, click “Next”. 11. Click “Finish” 12. Click “Next” Installation 4-12

Strona 18

13. Choose “I accept the license …….. “ and then click “Next” 14. Choose the default option “complete “ and then click “Next” Installation 4-13

Strona 19

15. Click “Next” 16. Click “Finish” Installation 4-14

Strona 20

17. Choose “Yes, I want to restart my computer now” and click “OK” Installation 4-15

Strona 21

4.4.4 Onyx-Rad Installation The Onyx-Rad software is supplied on a CD separate from the “Scanner Interface” CD. For installation of the Onyx-Rad sof

Strona 22 - Installation 4-10

4.4.5 Licensing the GOP Filters This section can be applied ONLY after installing the Dongle. This process is performed automatically with the scann

Strona 23

Safety Summary i Safety Summary ULIFTING HAZARD The PcCR 1417 scanner weights 40 Kg (88lb). • Do not try to lift the scanner by yourself. • Alway

Strona 24

5. The licensing window appears. Enter the code supplied by Orex into the “Key” dialog box. If you have not received the code, please contact your

Strona 25

Figure 9: CVLicense Window after GOP Licensing 7. The previous screen appears again. This time the “CVIE-Orex-Soft” on the 5ht row, the message “li

Strona 26

4.5 Installation Report For the warranty of the PcCR 1417 to take effect, and in order to verify correct installation by Orex, you are required to

Strona 27

Scanner Interface Software 5-1 5. Scanner Interface Software 5.1 General The Scanner Interface Software is a module that enables the FSE to access

Strona 28 - 4.4.4 Onyx-Rad Installation

Scanner Interface Software 5-2 Title Description 4. Erase For erasing the image on the phosphor plate. 5. Eject For unloading the plate; then ej

Strona 29

Scanner Interface Software 5-3 5.2.2 Setup Anatomical Note: Each setup relates to a specific part, to be scanned from a plate that originated fro

Strona 30

Scanner Interface Software 5-4 Title Parameter Description Organ Organ List Select the required organ from the drop-down list. The following organ

Strona 31

Scanner Interface Software 5-5 Title Parameter Description tune the image. If the image is too dark, increase the value by 10 numbers. If the imag

Strona 32 - 4.5 Installation Report

Scanner Interface Software 5-6 5.2.3 Setup Image Figure 12: Setup Menu, Image tab Title Parameter Description Jpg image properties Enable the o

Strona 33 - 5.2 Interface Description

Scanner Interface Software 5-7 Big Plate Resolution Always Ask Normal Resolution Before scanning a small plate, you will be prompted to choose

Strona 34

Laser Safety Instructions 1. During normal operation, to prevent the surronding area from being exposed to laser, the scanner should always be enclo

Strona 35 - 5.2.2 Setup Anatomical

Scanner Interface Software 5-8 5.2.4 Setup Configuration Figure 13: Setup Menu, Configuration tab Title Parameter Description Local Scanner Setti

Strona 36

Scanner Interface Software 5-9 Title Parameter Description Erase Factor Increase this value for a longer erasing process (When a very high dose w

Strona 37

Scanner Interface Software 5-10 Title Parameter Description Image Diagnostic For the calibration process. Once this option is checked, after the

Strona 38 - 5.2.3 Setup Image

Scanner Interface Software 5-11 Title Parameter Description files\orex\ACL4SB\archive Warming up override Option for technicians to disable the l

Strona 39

Scanner Interface Software 5-12 5.2.5 Setup Operation Update Figure 14: Setup Menu, Operation Update tab Title Parameter Description Unit 1-4 Back

Strona 40 - 5.2.4 Setup Configuration

Scanner Interface Software 5-13 5.2.6 Setup Diagnostics Figure 15: Setup Menu, Diagnostics tab Title Parameter Description Optical Erase Lamp Lamp

Strona 41

Scanner Interface Software 5-14 Title Parameter Description Plate Control Sensors Sensors Cassette in place – Cassette is docked Cassette hold –

Strona 42

Scanner Interface Software 5-15 5.2.7 Setup Automatic Tests This menu is used primarily, at the end of the Orex production line, for burn-in tests.

Strona 43

Scanner Interface Software 5-16 5.2.8 Setup Advance Figure 17: Setup Menu, Advance tab Title Parameter Description From here you can edit the seri

Strona 44

Scanner Interface Software 5-17 Title Parameter Description Save For saving the parameters. This HAS to be done before selecting another tab.

Strona 45 - 5.2.6 Setup Diagnostics

Table of Contents 1. Terms and Conventions...1-1 2. Packaging...

Strona 46

Scanner Interface Software 5-18 5.2.9 Setup About Figure 18: Setup Menu, About tab Title Parameter Description Software Version 1. Motion Curren

Strona 47 - 5.2.7 Setup Automatic Tests

Scanner Interface Software 5-19 Title Parameter Description Scanner ID Serial number Scan counter Scanner serial number Number of scans been done

Strona 48 - 5.2.8 Setup Advance

Scanner Interface Software 5-20 5.2.10 Image Diagnostics For performing final calibration of the filters assigned to the displayed part (sub-organ)

Strona 49

Scanner Interface Software 5-21 Title Parameter Description Data Correction method Data minimum For changing the data correction method. The recom

Strona 50 - 5.2.9 Setup About

Scanner Interface Software 5-22 It is possible to view the histogram of the image. In addition, for each sub-organ you can set a Data Minimum value.

Strona 51

Troubleshooting 6-1 6. Troubleshooting Problem What should I do? Remarks 1. “Disconnect” error message appears on the “scanner interface” scr

Strona 52 - 5.2.10 Image Diagnostics

7. Appendix A: Installation Prerequisites PRE-INSTALLATION FORM Computer requirements: PC: CPU – Pentium 4, 1.8GHz (min.) RAM – 256MB Hard-Driv

Strona 53

Appendix A: Installation Prerequisites 7-2 Single Media Archive required? Yes / No • When using the PcCr 1417 with a main PACS on site, the image

Strona 54

8. Appendix B: Installation Report INSTALLATION REPORT Reporting distributor: Company name: Technician installing the PcCr 1417: Name: Tee. No

Strona 55 - 6. Troubleshooting

5.2.1 Main Screen...5-1 5.2.2 Setup Anatomical ...

Strona 56 - Prerequisites

List of Figures Figure 1: Closed Box...2-

Strona 57

1. Terms and Conventions The following abbreviations are used throughout this manual: Term Meaning FSE Field Service Engineer PSP Photo-Stimulate P

Strona 58 - ** Optional

2. Packaging The PcCR 1417 is shipped in a custom designed box that protects the scanner. If the PcCR 1417 needs to be returned to Orex or to one of